| Catalog Number |
18061006 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
07/27/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The customer reported that:
"a 40-year-old patient was admitted to the hospital following a traffic accident with a left femoral diaphyseal fracture requiring.Description of the incident osteosynthesis using an intramedullary nail, model ref.1825-1140s.During the pre-disinfection of the nail holder after the nail was inserted, the operator noticed that a piece of the nail holder was missing and notified the surgeon.Clinical consequences and current status of the patient : no consequences for the patient; the procedure was able to proceed.Presence of a 3x3x1 mm metal fragment inside of the implanted nail.Very low risk of migration according to the surgeon.Any migration would be limited to the bony structures with no other associated risk.".
|
| |
|
Additional Manufacturer Narrative
|
|
Once the investigation has been completed any additional information will be reported in a supplemental report.
|
| |
|