| Model Number |
Mi1250 SYNCHRONY 2 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Tissue Breakdown (2681)
|
| Date of Event |
08/10/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Implant site is swollen, and a wound with infection and extrusion reported.
|
| |
|
Event or Problem Description
|
|
Implant site is swollen, and a wound with infection and extrusion reported.The device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: device investigation did not reveal any device defect or damage, which has been present whilst implanted.Mechanical damages found during investigation are attributable to the removal surgery.This finding was expected as the device was explanted due to swelling at the implant site leading to wound infection and extrusion.Review of the device_s sterilization records show that the device has been subject to a valid sterilization process, thus no available information points to the implant being the source of the reported issue.However, the presence of the device might have contributed to its subsequent development.This is a final report.
|
| |
|