MED-EL Elektromedizinische Geraete GmbH SYNCHRONY; Cochlear Implant Back to Search Results
Model Number Mi1250 SYNCHRONY 2
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Tissue Breakdown (2681)
Date of Event 08/10/2026
Type of Reportable Event Serious Injury
Event or Problem Description
Implant site is swollen, and a wound with infection and extrusion reported.
 
Event or Problem Description
Implant site is swollen, and a wound with infection and extrusion reported.The device was explanted.
 
Additional Manufacturer Narrative
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Additional Manufacturer Narrative
Conclusion: device investigation did not reveal any device defect or damage, which has been present whilst implanted.Mechanical damages found during investigation are attributable to the removal surgery.This finding was expected as the device was explanted due to swelling at the implant site leading to wound infection and extrusion.Review of the device_s sterilization records show that the device has been subject to a valid sterilization process, thus no available information points to the implant being the source of the reported issue.However, the presence of the device might have contributed to its subsequent development.This is a final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHRONY
Common Device Name
Cochlear Implant
Manufacturer (Section D)
MED-EL Elektromedizinische Geraete GmbH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
+43()577885626
MDR Report Key50544353
Report Number9710014-2026-00812
Device Sequence Number10318422
Product Code MCM
UDI-Device Identifier09008737369026
UDI-Public(01)09008737369026
Combination Product (Y/N)N
Initial Reporter CountryTC
PMA/510(K) Number
P000025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberMi1250 SYNCHRONY 2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/25/2026
Initial Report FDA Received Date09/01/2026
Date Device Manufactured12/06/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age28 MO
Patient SexMale
-
-