| Model Number |
HD Scorpion Needle |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
07/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2026, an fda medsun notification was received via email.A facility representative reported that an ar-13999hdn hd scorpion needle with megaloader had the tip of the needle break off and remain in the patient during the intraoperative phase.The surgeon confirmed that a 2.5 mm metal tip was retained in the rotator cuff.The surgeon stated that attempts to identify the retained tip were unsuccessful and that further disruption of the rotator cuff to locate the tip was not pursued because it would cause harm.The surgeon explained the finding to the patient and documented the discussion.The surgeon also discussed the finding with the radiologist, and it was determined that manipulation of the rotator cuff should be discontinued and that retention of the tip would not affect the clinical outcome.This was discovered during a left shoulder arthroscopy procedure in (b)(6) 2026.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|