The account reported finding ¿extension tubing keeps disconnecting,¿ and reported that the patient was on 2.5l of oxygen with multiple oxygen extension tubing connected together; the tubing disconnected twice, with saturations of 89% and 82%, no respiratory distress noted, and recovery with reconnection.A review of complaint history for the past 12 months showed no previous complaint with the product, no sample or photo was received for the investigation, trending review identified no trends, and the complaint was reviewed by both vendors.Due to not receiving the actual sample used, medline was not able to confirm what may have caused the issue.Based on the vendors response the root cause could not be determined, and the quality team will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|