| Model Number |
1020001 |
| Medical Device Problem Code |
Thermal Decomposition of Device (1071)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The manufacturer received a work order on an everflo device.During evaluation the technician noted front case melted.The technician noted that the faulty sieve canister assembly was the confirmed cause.Additionally unrelated to the reportable malfunction it was noted that a bad compressor caused low purity, and a faulty board caused the alarms not to be audible.In conclusion it was noted that the sieve canister, compressor, circuit board, front case, and the flow meter need to be replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported failure of thermal issue is accommodated in the product risk management file as being linked to hazard with description fire, flame, smoke.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no further indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.
dhr review was performed for the complaint device serial number: (b)(6) review confirms that the device met the final acceptance criteria and was released for final distribution.
|
| |
|
Search Alerts/Recalls
|
|
|