| Model Number |
97716 |
| Medical Device Problem Code |
Therapy Delivered to Incorrect Body Area (1508)
|
| Health Effect - Clinical Code |
Inappropriate Device Stimulation of Tissue (4928)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient noted that when stimulation was increased to obtain lower-extremity relief, intense stimulation or buzzing in the ribs was reported with all programs.Reprogramming was performed using both implanted leads, and rib stimulation continued.Some lower-extremity relief was reported when stimulation was maintained at a low level sufficient to avoid rib stimulation.An impedance check was performed, and all contacts were within normal limits.Thoracic-spine imaging was ordered, and follow-up was planned after imaging was completed.The stimulation-related rib sensation remained ongoing.The manufacturer representative (rep) indicated they would send any further information as they become aware.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|