• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Inogen, Inc. Inogen One G4 Bluetooth Portable Oxygen Concentrator

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Inogen, Inc. Inogen One G4 Bluetooth Portable Oxygen Concentrator Back to Search Results
Model Number IO-400
Medical Device Problem Codes Overheating of Device (1437); Unexpected Shutdown (4019)
Health Effect - Clinical Code Dyspnea (1816)
Date of Event 08/07/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient's device was getting hot and shutting off while going to the doctor.As a result, the patient experienced shortness of breath.No additional information is available at this time as multiple attempts to contact the patient were unsuccessful.
 
Additional Manufacturer Narrative
The inogen team attempted to contact the patient for further information three times with no response.As the device is not returned, no other information is available at this time to determine any other probable cause and/or the exact cause of the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Inogen One G4 Bluetooth Portable Oxygen Concentrator
Common Device Name
Inogen One G4 Bluetooth Portable Oxygen Concentrator
Manufacturer (Section D)
Inogen, Inc.
600 shiloh rd
plano TX 75074
Manufacturer (Section G)
Inogen, Inc.
600 shiloh rd
plano TX 75074
Manufacturer Contact
greg porter
600 shiloh rd
plano, TX 75074
MDR Report Key50547752
Report Number3004672275-2026-00278
Device Sequence Number10311939
Product Code CAW
UDI-Device Identifier00817131020179
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K032818
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberIO-400
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/13/2026
Initial Report FDA Received Date09/01/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/28/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening;
Patient SexUnknown
-
-