This complaint is recorded by zimmer biomet under (b)(4).G2: foreign - event occurred in united kingdom.Complaint can be confirmed through the returned product analysis.Review of the most recent repair record identified the following related repair: during pm it was noted that the device was out of calibration, width plate and machined head damaged.The machined head, pin, bearings, control bar, fine adj cam, semi-circle shaft bearing, vespel sleeve bearing and screws were replaced.Review of the device history record identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|