It was reported a cardiac perforation and pericardial effusion occurred.A left atrial appendage (laa) closure procedure was being performed.A watchman fxd curve access system (was) and a watchman laa closure device & delivery system (wds) were selected for use.During the initial transeptal puncture (tsp) using the was and versacross connect system, intracardiac echo (ice) was difficult to visualize the left atrium (la).The versacross connect wire was tented, and while preparing to buzz across the left atrium (la), it was noted the watchman fxd curve sheath and dilator had not advanced over to the la.The physician pulled back the wire to recross, noting again, the sheath and dilator had not advanced over the interatrial septum.At this time, it was noted the wire had perforated the aorta.Physician observed a small circumferential pericardial effusion, noting it was not large enough to drain/treat.Transesophageal echocardiography (tee) and computed tomography (ct) imaging was also completed as part of the assessment.The patient had stable blood pressure, and the implant procedure was discontinued.The patient remained in the hospital for three days.There were no further complications and the patient was discharged.
|