It was reported a pericardial effusion occurred.A left atrial appendage closure (laac) procedure was being performed.A 24mm watchman flx pro laa closure device & delivery system (wds) were selected for use.Transesophageal echocardiography (tee) imaging noted a trace effusion prior to the procedure.The initial transeptal puncture (tsp) was difficult to perform due to a sharp posterior-inferior bend in the patient anatomy.The physician attempted to cross the septum several times using the versacross connect wire, adding two different sheath/dilator bends for greater reach.After physician deployed the closure device, and completed a sweep, tee imaging showed no change in effusion.Approximately thirty minutes post deployment, the patient's blood pressure dropped.Tee imaging noted the effusion near the apex of the left ventricle.Physician performed a pericardiocentesis to treat the effusion, draining an unknown amount of dull colored fluid.Patient was stabilized and was admitted to intensive care unit.There were no further complications and approximately three days post procedure, the patient was discharged.
|