| Catalog Number |
e1651-88 |
| Medical Device Problem Codes |
Material Fragmentation (1261); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/28/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
We have determined that (b)(4) units from tuftex® over-the-wire embolectomy catheter lot number: xot00002181 were inadvertently sold to 11 us customers prior to being released.The lot was on hold for an ncmr investigation, where 1 out of 8 samples failed pull testing during manufacturing.It is possible that products from this lot are at a higher risk of experiencing ligature slip resulting in balloon deflation issues or balloon detachment during use.This failure could lead to adverse events during use.However, no complaints have been received related to these devices.Based on the issue, a field corrective action is being initiated by lemaitre to recall these devices (class ii recall).
|
| |
|
Additional Manufacturer Narrative
|
|
The (b)(4) units from lot: xot00002181 were sold to customers prior to the lot being released for sale.The lot was on hold under an ncmr for further investigation following 1 out of 8 samples failing the pull test during manufacturing with a ligature slip failure mode.It is possible that products from this lot are at a higher risk of experiencing ligature slip resulting in balloon deflation issues or balloon detachment during use.This failure could lead to adverse events during use.
|
| |
|