| Model Number |
M00558710 |
| Medical Device Problem Codes |
Burst Container or Vessel (1074); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to boston scientific corporation (bsc) that a cre pro wireguided dase dilatation balloon was attempted to be used in the esophagus during an esophageal stricture procedure performed on (b)(6) 2026.During the procedure, resistance was encountered when threading the balloon through the working channel of the scope.Upon inflation with the syringe assembly, the balloon burst inside the patients esophagus.The customer stated that no pieces of the balloon detached inside the patient.The procedure was completed with another cre pro wireguided dase dilatation balloon of a different lot number and a non-bsc inflation device.There were no patient complications reported as a result of this event.The patients condition at the conclusion of the procedure was reported to be stable.Additional information received 03 sep 2026.Patient was having an esophagogastroduodenoscopy (egd) dilation and a cre pro wireguided dilator measurements 18-20mm was used.Instrument appeared intact, all checks completed.The balloon was inflated as normal, but the balloon burst while in the patient's esophagus at the max pressure.No injury to the patient, however, a new instrument had to be used during the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6:
imdrf device code a0402 captures the reportable event of balloon burst.Block h11: investigation results:
the device was not returned for analysis and was reported to be disposed; therefore, a technical analysis could not be performed.With all available information, boston scientific concludes the reported event of balloon burst could not be confirmed.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.The device history record review confirmed that the device met all material, assembly, and performance specifications prior to distribution.A similar complaint review was completed.There were no emerging trends identified that warrant further action.
|
| |
|
Search Alerts/Recalls
|
|
|