• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. EVIS EXERA II COLONOVIDEOSCOPE; Colonoscope And Accessories, Flexible/Rigid

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. EVIS EXERA II COLONOVIDEOSCOPE; Colonoscope And Accessories, Flexible/Rigid Back to Search Results
Model Number CF-H180AI
Medical Device Problem Codes Erratic or Intermittent Display (1182); Poor Quality Image (1408)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/31/2026
Type of Reportable Event Malfunction
Event or Problem Description
No additional information received from the customer.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to provide the results of the final investigation.Updated fields: d8, h2, h6, h11.The device was returned to olympus for inspection, the reported failure and the failure identified during initial evaluation were confirmed.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: damage on switch.The cause is traced to a component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II COLONOVIDEOSCOPE
Common Device Name
Colonoscope And Accessories, Flexible/Rigid
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
+1(508)207-7661
MDR Report Key50550202
Report Number9610595-2026-71732
Device Sequence Number16101373
Product Code FDF
UDI-Device Identifier04953170339585
UDI-Public04953170339585
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K100584
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Type of Report Initial,Followup
Report Date (Section B) 09/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCF-H180AI
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/14/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/11/2026
Supplement Date Received by Manufacturer09/01/2026
Initial Report FDA Received Date09/01/2026
Supplement Report FDA Received Date09/28/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/24/2008
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-