The account reported tbg insufflation suct whtie (insufflation tubing) was not fitting into wall suction and the end of the tubing is not compatible with their devices or wall hook ups.A photo was received for the investigation; the photo shows only a white tip, but no tubing, finished good production records were reviewed with no non-conformances noted, and the finished good pack build was reviewed with no issues that might have an impact on the issue.After further review, the photo of the white tip the account sent in is not in this kit's bill of materials, and to confirm this is an assembly issue, the full tubing component would be needed.The cause code was adjusted to unconfirmed defect with item not in kit bom, and the division will monitor for trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|