| Model Number |
A22040A |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/25/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
No additional information received from customer.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the final investigation.Update fields: g3, h2, h3, h6 and h11.The device was returned to olympus for inspection.Based on the results of the investigation the customer allegation was confirmed that the ceramic head (ceramic tip/insulation beak) was missing during the service inspection.It was likely due to stress problem.The most probable cause of this complaint is component failure.The root cause could not be determined.Olympus will continue to monitor field performance for this device.
|
| |
|