Medtronic received information regarding a navigation system being used during a sacroiliac and thoracolumbar procedure.It was reported that there was an alleged inaccuracy.During an oblique lumbar interbody fusion (olif) on l2-l3, they used the left side of the navigation system to take a navigated spin and the screw went into the canal on l2 which may have paralyzed the patient.When they took the screw out, the patient reportedly "almost jumped off the table." they switched to an open posterior procedure and re-spun using the right imaging system trackers and reported everything appeared to be accurate.This issue caused a 45-minute surgical delay.No additional information regarding the patient's outcome was reported.
|
|
Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: 9735740, software version: 2.1.0 h3,h6: no products have been returned to medtronic for analysis.Codes b20, c20, and d15 are applicable.H6: code f24: insufficient health impact information is applicable to no additional information regarding the patient's outcome was reported.Code f1908: prolonged surgery is applicable to the 45-minute surgical delay.Code f1906: modified surgical procedure is applicable to they took the screw out, and they switched to an open posterior procedure.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|