| Model Number |
419588 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); High impedance (1291); Over-Sensing (1438); Pacing Problem (1439)
|
| Health Effect - Clinical Codes |
Dyspnea (1816); Palpitations (2467)
|
| Date of Event |
06/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the right atrial (ra) lead exhibited out of range (oor) high pacing impedance , oversensing resulting in inhibition of pacing and misclassification of episode type.The lead also exhibited noise. the patient experienced palpitations and shortness of breath.The lead remains in use. no patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Continuation of d10: w1dr01 ipg - implanted (b)(6) 2026.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|