The customer reported that the gloves were "not the same quality as all the previous gloves we have received," with texture "thin," "very brittle," and "tearing easily," and the box had "a lighter color throughout the box." the account provided one used box; thirteen gloves were donned and evaluated, including measurements of glove thickness, palm width, and overall length from the tip of the middle finger to the cuff, and no gloves tore during the donning and doffing testing while all measured dimensions remained within specification.Based on the testing performed and the measurement results, medline was unable to confirm the reported issue, and the root cause could not be identified.There were no reports of death or serious injury associated with this event, and there has been no change to the reportability decision associated with this complaint.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|