It was reported that during implantation of a single-piece non-preloaded monofocal intraocular lens (iol), the capsule broke.As a result, the doctor did not implant the reported lens and opted to use a ar40e lens model to complete the procedure.The event occurred during implantation/application.The product was not available for return for investigation.No further information was provided.
|
|
Section a2: date of birth, age: unknown/not provided
section a3a, a3b: sex, gender: unknown/not provided
section a4: weight: unknown/not provided
section a5: ethnicity: unknown/not provided
section a6: race: unknown/not provided
section d6a: if implanted, give date: unknown/not provided; the extent of patient contact is unknown, however there is no indication that the lens was fully implanted.Section d6b: if explanted, give date: unknown/not provided; the extent of patient contact is unknown, however there is no indication that the lens was fully implanted, therefore, not explanted.Device evaluation:
product evaluation was not performed because the product has not been received.The complaint issue reported could not be confirmed.Manufacturing record review:
the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.A search in the complaint system revealed that no other complaints were received for this production order number.Conclusion:
based on the investigation, no malfunction or product deficiency was identified.Attempts were made to obtain the missing information; however, no additional information has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|