| Model Number |
ZA9003 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
02/02/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a fully implanted three-piece spherical monofocal intraocular lens (iol) was found to be dislocated on the left eye.The patient symptoms was noted to have persisted for approximately 7.5 years.The issue was observed during a postoperative examination conducted during a site visit.The original lens was subsequently explanted during a secondary surgery and was replaced with a ar40e monofocal intraocular lens (iol) with a +22.0d.No unplanned surgical interventions such as, incision enlargement, sutures, or vitrectomy were required.Directions for use was followed.The patient has fully recovered.The patient was fully recovered.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation:
product evaluation was not performed because the product has not been received.The complaint issue reported could not be confirmed.Manufacturing record review:
the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.A search in the complaint system revealed that no other complaints were received for this production order number.Conclusion:
based on the investigation, no malfunction or product deficiency was identified.Attempts were made to obtain the missing information; however, no additional information has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|