• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO Puerto Rico Mfg. Inc. TECNIS SIMPLICITY; Intraocular Lens

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO Puerto Rico Mfg. Inc. TECNIS SIMPLICITY; Intraocular Lens Back to Search Results
Model Number DIB00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Capsular Bag Tear (2639)
Date of Event 08/05/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a preloaded monofocal intraocular lens (iol) was fully implanted and subsequently removed during the same surgical procedure due to a posterior capsule rupture.A vitrectomy and sutures were required.The procedure was completed successfully using a replacement non-johnson & johnson lens (model: ma60ac, serial number: (b)(6)).The lens and cartridge came into contact with the patient's left eye.The physician indicated that the device did not cause or contribute to the event.The patient's outcome was reported as fine.Additional information received confirmed that the posterior capsule rupture occurred after implantation of the intraocular lens (iol).The lens was successfully injected into the capsular bag, and the trailing haptic was positioned using a sinskey hook.During removal of the residual viscoelastic with an irrigation/aspiration device, a posterior capsule rupture occurred.The physician did not identify any issues with the lens, cartridge, or implantation process that may have contributed to the event.Furthermore, no preexisting patient conditions were identified as contributing factors.As a result of the posterior capsule rupture, an unplanned anterior vitrectomy was performed, and sutures were required.No other unplanned surgical interventions, such as incision enlargement or additional procedures, were necessary, and no medication outside the standard of care was prescribed.The implanted lens was removed as a direct consequence of the posterior capsule rupture.The patient's current clinical status and visual outcome were reported as stable.No patient injuries or postoperative complications related to the event were reported.The postoperative visual acuity and refraction were reported as 20/30 without glasses.No additional information was provided.
 
Additional Manufacturer Narrative
Section a4: the weight of 99.3 kilograms did not populate in the field because the system does not accept decimal values.Section d6a: not applicable, as the lens was inserted and removed in the same procedure.Section d6b: not applicable, as the lens was inserted and removed in the same procedure.Section h3:the device was not returned for evaluation.Therefore, a failure analysis of the complaint device cannot be completed.A review of records including device history and complaint trending will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Section h6: health effect-impact code:4625-additional surgery (sutures and unplanned vitrectomy).Attempts has been made to obtain missing information; however, the information has not been provided.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SIMPLICITY
Common Device Name
Intraocular Lens
Manufacturer (Section D)
AMO Puerto Rico Mfg. Inc.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
+1(714)247-8552
MDR Report Key50554029
Report Number3012236936-2026-000256
Device Sequence Number10601146
Product Code HQL
UDI-Device Identifier05050474731714
UDI-Public(01)05050474731714(17)290105
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberDIB00
Device Catalogue NumberDIB00U0180
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/24/2026
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/06/2026
Supplement Date Received by Manufacturer09/02/2026
Initial Report FDA Received Date09/01/2026
Supplement Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/05/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age75 YR
Patient SexFemale
Patient RaceWhite
-
-