• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR Surgical Company EVO/EVO + VISIAN Implantable Collamer Lens (ICL); Phakic Toric Intraocular Lens

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR Surgical Company EVO/EVO + VISIAN Implantable Collamer Lens (ICL); Phakic Toric Intraocular Lens Back to Search Results
Model Number VTICM5 13.2 (-13.5/ 1.0)
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Intraocular Pressure Increased (1937)
Date of Event 08/23/2026
Type of Reportable Event Serious Injury
Event or Problem Description
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced an excessive vault and elevated iop.Medication was prescribed.In a separate surgery the lens was explanted.The cause of the event was reported as device/user error/ unknown.Cause details provided: 'we think about wtw'.There are multiple medical device reports associated with this patient.This report is 1 of 2 total reports.Based on the information provided, the risk assessment for our product, and our experience, we have determined the cause or contribution to the reported issue is not indicative of a manufacturing related issue or product deficiency.
 
Additional Manufacturer Narrative
H6- investigation type code: 4110- lens work order search-no similar complaint event(s) within associated lots were found.H11- manufacturer narrative: iop elevation from baseline and secondary surgical intervention to remove/replace/reposition the len is identified in the labeling as a known potential adverse event following icl implantation.Claim# 749084.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVO/EVO + VISIAN Implantable Collamer Lens (ICL)
Common Device Name
Phakic Toric Intraocular Lens
Manufacturer (Section D)
STAAR Surgical Company
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR Surgical Company
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene enriquez
1911 walker avenue
monrovia, CA 91016
MDR Report Key50554770
Report Number2023826-2026-01665
Device Sequence Number10651038
Product Code QCB
UDI-Device Identifier00840311311255
Combination Product (Y/N)N
Initial Reporter CountryIZ
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberVTICM5 13.2 (-13.5/ 1.0)
Device Catalogue NumberVTICM5 13.2
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/25/2026
Initial Report FDA Received Date09/02/2026
Date Device Manufactured11/03/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Cartridge Model: SFC-45: Lot# UNK; Foam Tip Plunger (FTP), Lot# UNK; Injector Model: MSI-PF, Lot# UNK
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-