Datascope Corp. Sensation Plus 8Fr. 50cc IAB with Accessories, US Only; System, Balloon, Intra-Aortic And Control Back to Search Results
Model Number 0684-00-0576-01U
Medical Device Problem Code Restricted Flow rate (1248)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/13/2026
Type of Reportable Event Malfunction
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
Please revert the a1 (patient id) and d8, to blank updated fields: b1,b4,g3,g6,g7,h2,h3,h6 (type of investigation, investigation findings, component codes, investigation conclusions), h11 corrected fields: g4.The iab was returned with the membrane completely unfolded and blood on the exterior of the catheter and between the catheter and the sheath.The returned sheath was over the catheter and partially covering the rear portion of the balloon.The extender and pressure tubing was also returned.An underwater leak test of the balloon, catheter, y fitting, extracorporeal, pressure and extender tubing was performed and no leaks were detected.The iab was placed on the cardiosave pump and the iab fully inflated and deflated.The iab then pumped for two hours which represents one complete autofill cycle.The iab pumped normally and no alarm sounded from the pump.The reported problem cannot be confirmed by the evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Sensation Plus 8Fr. 50cc IAB with Accessories, US Only
Common Device Name
System, Balloon, Intra-Aortic And Control
Manufacturer (Section D)
Datascope Corp.
15 law drive
fairfield NJ 07004
Manufacturer (Section G)
Datascope Corp.
15 law drive
fairfield NJ 07004
Manufacturer Contact
ewelina chorlowska
15 law drive
fairfield, NJ 07004
MDR Report Key50558250
Report Number2248146-2026-0001493
Device Sequence Number16085418
Product Code DSP
UDI-Device Identifier10607567109633
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K112327
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 09/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0684-00-0576-01U
Device Catalogue Number0684-00-0576-01U
Device Lot Number3000507487
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/28/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/13/2026
Initial Report FDA Received Date09/02/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/16/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient Age64 YR
Patient SexMale
Patient Weight193 LB
-
-