Please revert the a1 (patient id) and d8, to blank
updated fields: b1,b4,g3,g6,g7,h2,h3,h6 (type of investigation, investigation findings, component codes, investigation conclusions), h11
corrected fields: g4.The iab was returned with the membrane completely unfolded and blood on the exterior of the catheter and between the catheter and the sheath.The returned sheath was over the catheter and partially covering the rear portion of the balloon.The extender and pressure tubing was also returned.An underwater leak test of the balloon, catheter, y fitting, extracorporeal, pressure and extender tubing was performed and no leaks were detected.The iab was placed on the cardiosave pump and the iab fully inflated and deflated.The iab then pumped for two hours which represents one complete autofill cycle.The iab pumped normally and no alarm sounded from the pump.The reported problem cannot be confirmed by the evaluation.
|