H3: product analysis confirmed 'making strange noise.' unit presented with misaligned crm radio firmware, and a defective eic pcba.H6 : the previously applied codes are b17, c20, d16 and img g02030 are no longer applicable.Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: nim4cpb1, serial/lot #: (b)(6), ubd: , udi#: (b)(4) h3: product analysis confirmed 'making strange noise.' unit presented with sw:(v1.7.5/field update), misaligned radio firmware, fully charged batteries, a worn fuse decal, and a defective afe pcba.H6 : the previously applied codes are b17, c20, d16 and img g02030 are no longer applicable.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|