Model Number NIM4CM01
Medical Device Problem Code Inappropriate Audible Prompt/Feedback (2280)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/31/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3: product analysis confirmed 'making strange noise.' unit presented with misaligned crm radio firmware, and a defective eic pcba.H6 : the previously applied codes are b17, c20, d16 and img g02030 are no longer applicable.Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: nim4cpb1, serial/lot #: (b)(6), ubd: , udi#: (b)(4) h3: product analysis confirmed 'making strange noise.' unit presented with sw:(v1.7.5/field update), misaligned radio firmware, fully charged batteries, a worn fuse decal, and a defective afe pcba.H6 : the previously applied codes are b17, c20, d16 and img g02030 are no longer applicable.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIM 4.0 CONSOLE
Common Device Name
STIMULATOR, ELECTRICAL, EVOKED RESPONSE
Manufacturer (Section D)
Medtronic Xomed Inc.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer (Section G)
Medtronic Xomed Inc.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer Contact
justin ellis
6743 southpoint drive north
jacksonville, FL 32216
+1(763)526-5677
MDR Report Key50558736
Report Number1045254-2026-01217
Device Sequence Number10533456
Product Code GWF
UDI-Device Identifier00763000395896
UDI-Public(01)00763000395896(11)230119
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K200759
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberNIM4CM01
Device Catalogue NumberNIM4CM01
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/14/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/10/2026
Supplement Date Received by Manufacturer09/14/2026
Initial Report FDA Received Date09/02/2026
Supplement Report FDA Received Date10/01/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/19/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-