| Catalog Number |
SK15018 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/11/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2026, a patient underwent a recanalization procedure using a seeker support catheter via femoral access.During the procedure, it was reported that the catheter ruptured while being advanced over a 0.018-inch guidewire.The catheter separated into two pieces.The procedure was subsequently completed using a competitor's catheter.No patient harm was reported, and there was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: this supplemental report is submitted to correct an error in the previously submitted reports.The initial mdr and supplemental report 2 listed an incorrect manufacturing location (g1).The correct manufacturing location for this device is futurematrix interventional.No new event information is being reported in this supplement.Manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this lot.Investigation summary: the physical device was not returned for evaluation.Two photos were provided and reviewed.First photo shows that the device seeker is been holding by the hcp in the detached condition.Second photo shows that the device and labelling details.Verified the same in the tw.Based on the photo review, the reported event can be confirmed.A definitive root cause for the detachment could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.G1, g3.Section a through f: the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Search Alerts/Recalls
|