| Model Number |
VH-3010 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/29/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The hospital reported that the t.W.Power supply was not working correctly during the procedure.The problem that occurred was that the device was not cauterizing.Power setting at 3.There was no change in the power settings, and no adjustment to the settings resolved the issue.The power supply was placed on a shelf near the bovie machine that was not on.The extension cable, adapter, and new harvesting tool kits were changed out; however, the issue persisted and the attempts were unsuccessful.A new device was used to complete the procedure, there was no patient harm.There was a small procedural delay, as the harvest time was lengthened to obtain another device.The power supply is a 2023 unit there is no way of knowing how old their cables and adaptors are because they don't have serial numbers on them.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Corrected sections: b5,
updated section: b4, e1 (phone #), g3, g6, h2, h11.
|
| |
|