H11: manufacturing review: the device history records have been reviewed, and this lot met all release criteria.Investigation summary: although the physical device was not returned for evaluation, one electronic photo was provided for review.The photo contains native language text boxes and shows a partial view of clinician holding drainage catheter in which crack was observed on the white catheter hub.Therefore, the investigation is confirmed for reported fracture issue.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.E1 (b)(6).B5, g3, h6 (method, result, conclusion).Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|