| Model Number |
Mi1200 SYNCHRONY |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/26/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The user noticed a change in hearing performance with the device.The user was reimplanted.
|
| |
|
Event or Problem Description
|
|
The user noticed a change in hearing performance with the device.The user was reimplanted.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: damage to the active electrode, likely caused by minute device mobility, was determined to have led to device failure over time.Other mechanical damages found are attributable to the removal surgery.The problems given in the recipient report appear to match the damage found.This is a final report.
|
| |
|
Search Alerts/Recalls
|