On (b)(6) 2026, a patient underwent a kidney biopsy procedure using a maxcore instrument.During the procedure, after two tissue samples were collected, the healthcare professional triggered the device to obtain a third sample.However, the outer sheath failed to deploy, and only the inner needle fired.The procedure was completed using another device.There was no reported patient injury.
|
|
H11: manufacturing review: a manufacturing review was not requested as the lot number reported is unknown.Investigation summary: received one 18g maxcore disposable core biopsy instrument.The device was received without a needle protector and without a yellow guard attached to the needle.The device appeared to have residue throughout was received in the initial position.Upon visual evaluation, no visual anomalies were noted to the device.On functional testing, the device was fully primed with no issues.The device was further tested by cocking and firing the device 3 times into the chicken breast using each trigger for a total of 6 tests.The device was able to prime and fire properly.All 6 samples were collected with no issue.Therefore, the investigation is unconfirmed for the reported failure as the device prime, fire properly and obtain all samples with no issue.A definitive root cause for the alleged failure to fire issue could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.E1, g3, h6 (method, result, conclusion).Section a through f: the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|