MAUDE Adverse Event Report: LivaNova Deutschland Electronic Gas Blender; Gas Control Unit, Cardiopulmonary Bypass

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LivaNova Deutschland Electronic Gas Blender; Gas Control Unit, Cardiopulmonary Bypass Back to Search Results
Model Number Electronic Gas Blender
Medical Device Problem Code Gas Output Problem (1266)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/07/2026
Type of Reportable Event Malfunction
Event or Problem Description
See initial report.
 
Additional Manufacturer Narrative
H11: through followup communications, it was learnt that event occurred while preparing an hlm system.Unit was returned to manufacturer: technical safety inspection including a prolonged stress test was conducted and device was found to work according to manufacturer specifications.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Electronic Gas Blender
Common Device Name
Gas Control Unit, Cardiopulmonary Bypass
Manufacturer (Section D)
LivaNova Deutschland
lindberghstraße 25
munich 80939
GM  80939
Manufacturer (Section G)
LivaNova Deutschland
lindberghstraße 25
munich 80939
GM   80939
Manufacturer Contact
annarita di lascio
14401 w. 65th way
arvada, CO 80004
MDR Report Key50565110
Report Number9611109-2026-90407
Device Sequence Number15977020
Product Code DTX
UDI-Device Identifier04033817900696
UDI-Public(01)04033817900696(11)250620
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K101046
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Type of Report Initial,Followup
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberElectronic Gas Blender
Device Catalogue Number25-40-45
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/07/2026
Supplement Date Received by Manufacturer08/31/2026
Initial Report FDA Received Date09/02/2026
Supplement Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/18/2024
Patient Sequence Number1