| Lot Number |
0039159150 |
| Medical Device Problem Codes |
Nonstandard Device (1420); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/05/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During an atrial fibrillation ablation, a farawave ablation catheter was selected for use.During preparation, it was noted that the wire had a slight resistance when pulling back into the catheter.After an additional inspection, a small fragmentation was noticed at the distal tip of the catheter.The catheter was replaced and procedure was completed with no patient complications.
|
| |
|
Event or Problem Description
|
|
During an atrial fibrillation ablation, a farawave ablation catheter was selected for use.During preparation, it was noted that the wire had a slight resistance when pulling back into the catheter.After an additional inspection, a small fragmentation was noticed at the distal tip of the catheter.The catheter was replaced and procedure was completed with no patient complications.
|
| |
|
Additional Manufacturer Narrative
|
|
D2a: common device name: percutaneous cardiac ablation catheter for treatment of atrial fibrillation with irreversible electroporation; reported here as the common device name exceeded the character limit for designated field.
|
| |
|
Additional Manufacturer Narrative
|
|
D2a: common device name: percutaneous cardiac ablation catheter for treatment of atrial fibrillation with irreversible electroporation; reported here as the common device name exceeded the character limit for designated field.Device technical analysis:
during product analysis the device passed all tests performed, showing no evidence of the reported failure.Additionally, no damage or defects were observed during the analysis.Device history record review:
it was confirmed this device met specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Labeling review:
based on the information provided, there is no evidence that the device was used in a manner inconsistent with the labelled indications/instructions for use (ifu).Risk review:
a risk review performed for the fawave device confirmed that the event of damage and difficult to load wire are known events defined in the products risk management documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Investigation conclusion:
boston scientific has assigned an investigation conclusion code of no problem detected for the reported allegations as the allegations were unable to be confirmed, and no issues were observed during the analysis.A guidewire was successfully inserted through the catheter.Deployment was tested multiple times, and the device was deployed to basket and flower position with no unusual resistance noted.The device passed all tests performed, showing no evidence of the reported failure.Additionally, no damage or defects were observed during the analysis.
|
| |
|