H11: manufacturing review: the device history records have been reviewed, and this lot met all release criteria.Investigation summary: although the sample was not returned for evaluation, one electronic photo and two videos were provided for review.The first video shows partial view of a clinician holding the drainage catheter.Fluid leak was noted connection between hub and catheter.The second video shows partial view of a clinician holding two implanted drainage catheters.Fluid leakage was observed at the junction between the catheter hub and the catheter in both drainage catheters.The photo shows the partial view of clinician holding three drainage catheter packaging.The product label along with a catheter was partially visible for one packaging.Therefore, the investigation is confirmed for the reported fluid leak issue for this first device.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.E1: (b)(6).G3, h6 (component, method, result, conclusion).Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|