| Model Number |
S46-937 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
08/06/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Technomed received a complaint via distributor consolidated neuro supply regarding a subdermal needle electrode used for posterior tibial nerve (ptn) stimulation.During removal of the needle from the patient's ankle, the needle reportedly broke, leaving a portion retained in the patient.An x-ray was performed and confirmed that the needle remained in the patient's ankle.Attempts to remove the retained portion using a magnet and hemostats were unsuccessful.It was subsequently reported that the patient was in interventional radiology for removal of the retained needle.The affected product was identified as ref.S46-937, lot 104474.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal investigation found no anomalies or product defects in retained samples.The event was considered consistent with the product's mechanical capability being exceeded rather than a systematic product issue.
|
| |
|
Event or Problem Description
|
|
Technomed received a complaint via distributor consolidated neuro supply regarding a subdermal needle electrode used for posterior tibial nerve (ptn) stimulation.During removal of the needle from the patient's ankle, the needle reportedly broke, leaving a portion retained in the patient.An x-ray was performed and confirmed that the needle remained in the patient's ankle.Attempts to remove the retained portion using a magnet and hemostats were unsuccessful.It was subsequently reported that the patient was in interventional radiology for removal of the retained needle.The affected product was identified as ref.S46-937, lot 104474.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer's investigation has been completed.The device was not returned for evaluation as it had been disposed of by the hospital.Evaluation of retained samples and review of the manufacturing and traceability records identified no abnormalities or relevant deviations.A definitive root cause could not be established.No evidence of a systematic product or manufacturing issue was identified.
|
| |
|