| Model Number |
97810 |
| Medical Device Problem Codes |
Charging Problem (2892); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Codes |
Discomfort (2330); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/01/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Information was received from a patient who was implanted with an implantable neurostimulator (ins) for gastrointestinal/pelvic floor, urinary dysfunction/sacral nerve stim.It was reported that the patient reported for the past 5 months they felt uncomfortable since the ins was not working because their wireless recharger was not charging and was not able to recharge their ins.The patient mentioned they saw the scrolling battery lights on the recharger and it was not emitting any tones.The patient added their doctor redirected them to call the manufacturer for assistance.The agent reviewed how to attempt a hard reset, but did not resolved the issue.The agent sent repair request.On 2026-jul-13 additional information was received from the patient.The patient stated that the replacement recharger resolved their issues.
|
| |
|
Additional Manufacturer Narrative
|
|
Continuation of d10: product id wr9220 lot# serial# (b)(6): product type recharger section d information references the main component of the system.Other relevant device(s) are: product id: wr9220, serial/lot #: (b)(6): h3: analysis of the wireless recharger (s/n (b)(6)) revealed that the patient complaint was verified.Leds scrolled continuously.The device was unresponsive.Flash sector read/write protection bits had become set.No anomalies were visually observed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|