| Model Number |
G124 |
| Medical Device Problem Code |
Unexpected Therapeutic Results (1631)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/11/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) delivered a shock, despite the patient only in ventricular tachycardia (vt).The device charged and diverted a shock as the rate went below the therapy threshold.The patient rhythm then reentered vt and the shock was delivered as the device cannot divert twice.Technical services (ts) confirmed the device operated per design.This device was subsequently explanted and returned.No additional adverse patient effects were reported.
|
| |
|