• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Masimo - 15750 Alton Pkwy Radius VSM; Cardiac Monitor (Including Cardiotachometer and Rate Alarm)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Masimo - 15750 Alton Pkwy Radius VSM; Cardiac Monitor (Including Cardiotachometer and Rate Alarm) Back to Search Results
Model Number 9037
Medical Device Problem Codes Incorrect Measurement (1383); Low Readings (2460)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/22/2026
Type of Reportable Event Malfunction
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Radius VSM
Common Device Name
Cardiac Monitor (Including Cardiotachometer and Rate Alarm)
Manufacturer (Section D)
Masimo - 15750 Alton Pkwy
15750 alton pkwy
irvine CA 92618
Manufacturer (Section G)
Masimo - Mexicali
industrial vallera de mexicali calzada del oro, no.2001
mexicali, baja california 21600
MX   21600
Manufacturer Contact
lusine boyadzhyan
15750 alton pkwy
irvine, CA 92618
+1(949)297-7167
MDR Report Key50568958
Report Number3019388613-2026-00593
Device Sequence Number15934652
Product Code MHX
UDI-Device Identifier00843997019163
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K223498
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 08/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9037
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/07/2026
Initial Report FDA Received Date09/02/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-