| Model Number |
JP2110X16B |
| Medical Device Problem Codes |
Thermal Decomposition of Device (1071); Circuit Failure (1089); Power Problem (3010)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/05/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A trilogy evo ventilator was returned to the manufacturer service center for preventative maintenance.The device was not in use on a patient at the time of the occurrence.During the evaluation the ac power was not recognized therefore the power supply was replaced to resolve the issue.During the final test thermal damage was observed on the ac power cord therefore, the system board was replaced to address the issue.Additionally, during the evaluation the device failed a fio2 sensor receptacle verification test step during testing unrelated to the identified malfunction, therefore, the inline proximal filter was replaced to address the issue.Periodic maintenance 2 was performed.The following parts were replaced to prevent failure: air inlet foam filter muffler assembly and particulate filter.The technician performed final test, battery check, valve check, run in tests and cleaning then confirmed the unit operated properly.The software version was upgraded from 1.06.10.00 to 1.06.15.00.
|
| |
|
Additional Manufacturer Narrative
|
|
Dhr review was performed for the complaint device sn.Review confirms that the device met the final acceptance criteria and was released for final distribution.The reported failure of thermal issue is accommodated in the product risk management file as being linked to hazard with description fire, flame, smoke.Complaints for this failure were reviewed to assess for the observed probability levels and compare with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.
|
| |
|
Search Alerts/Recalls
|
|
|