• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medtronic Puerto Rico Villalba Sprint Quattro MRI SureScan; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medtronic Puerto Rico Villalba Sprint Quattro MRI SureScan; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number 6946M62
Medical Device Problem Codes Signal Artifact/Noise (1036); Fracture (1260); Over-Sensing (1438); Impedance Problem (2950)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/31/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the right ventricular (rv) lead had a suspected fracture on the low voltage portion.There were warnings for oversensing noise and pacing impedance, and a lead integrity alert (lia) was triggered for the rv lead.Detection and alarms were turned off, and the lead remains in use.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Sprint Quattro MRI SureScan
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
Medtronic Puerto Rico Villalba
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
Medtronic Puerto Rico Villalba
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112--439
+1(763)526-9277
MDR Report Key50569820
Report Number2649622-2026-26117
Device Sequence Number10507415
Product Code LWS
UDI-Device Identifier00643169858329
UDI-Public(01)00643169858329(17)210411
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
P920015
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/11/2021
Device Model Number6946M62
Device Catalogue Number6946M62
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/14/2026
Supplement Date Received by Manufacturer09/08/2026
Initial Report FDA Received Date09/02/2026
Supplement Report FDA Received Date09/17/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/03/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5076-52 LEAD, DDMC3D4 ICD
Outcome Attributed to Adverse Event Required Intervention;
Patient Age76 YR
Patient SexMale
Patient Weight150 LB
-
-