| Model Number |
RFIT-ASY-0118 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/30/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Summary:
the customer (b)(6) reported a potential false positive varicella zoster virus (vzv) result on the filmarray® meningitis/encephalitis (me) panel after testing a patient csf sample.The customer reported that due to the me panel result, that patient was indirectly impacted, specifically initiation of acyclovir that was likely unnecessary, which might lead to serious injury if it were to recur.No patient harm was reported.The investigation concluded that the most likely cause of the vzv discrepant result was due to low level contamination.
|
| |
|
Event or Problem Description
|
|
The customer (dunedin, new zealand) reported a potential false positive varicella zoster virus (vzv) result on the filmarray® meningitis/encephalitis (me) panel after testing a patient csf sample.The customer reported that due to the me panel result, that patient was indirectly impacted, specifically initiation of antibiotic therapy that was likely unnecessary, which might lead to serious injury if it were to recur.No patient harm was reported.Biofire has requested further information from the customer and is currently investigating this event.No remedial action, corrective action, preventive action, or field safety corrective action (fsca) has been deemed necessary at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation:
the patient was a 1-day-old male preterm infant with increasing respiratory support needs.On (b)(6) 2026, the patient csf sample was collected via puncture and tested on the biofire filmarray meningitis/encephalitis (me) panel.The me panel reported varicella-zoster virus (vzv) as detected.The me panel was discordant with the clinical picture.On an unknown date, roche lightmix hsv/vzv pcr was performed on the same sample and yielded a negative result for vzv.The me panel was discordant with comparator methods.The customer reported that due to the me panel result, that patient was indirectly impacted, specifically initiation of antibiotic therapy (acyclovir) that was likely unnecessary, which might lead to serious injury if it were to recur.No patient harm was reported.The analysis of the me panel run file provided showed by the customer showed a late signature suggesting that the target nucleic acid was present at low levels.Quality control (qc) records for me panel pouch lot# 3va625 (kit lot# 1696125) and the instrument (serial number# 2fa00007) were reviewed.The pouch lot and instrument passed qc criteria.The discrepancy reported by the customer was not observed during qc testing.Conclusion:
the investigation concluded that the most likely cause of the vzv discrepant result was due to low level contamination.Due to the sensitive nature of the me panel, there is a risk of false positive results from the introduction of exogenous organism or nucleic acid.While the pouch loading instructions and recommended cleaning steps are sufficient to eliminate contamination the majority of the time, it is possible that some organism/nucleic acid could be introduced during any part of the testing and pouch handling process.Contamination can also be introduced via the laboratory environment and equipment such as pipettes, pipette tips, etc., or personnel handling the sample.Clinical performance can be found in tables 9 and 14 of the biofire® me panel ifu (https://www.Biofiredx.Qarad.Eifu.Online/iti/all?keycode=iti0035).
|
| |
|
Search Alerts/Recalls
|
|
|