Abbott Medical (CRM-Sylmar) Assert-IQ¿ ICM; Recorder, Event, Implantable Cardiac, (Without Arrhythmia Detection) Back to Search Results
Model Number DM5000
Medical Device Problem Code Over-Sensing (1438)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/13/2026
Type of Reportable Event Malfunction
Event or Problem Description
Additional information received indicated the oversensing episodes were due to post-sensed t-wave oversensing and far p-wave oversensing.The device was reprogrammed to resolve the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Assert-IQ¿ ICM
Common Device Name
Recorder, Event, Implantable Cardiac, (Without Arrhythmia Detection)
Manufacturer (Section D)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
Abbott
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key50570691
Report Number2017865-2026-18493
Device Sequence Number21381490
Product Code MXC
UDI-Device Identifier05415067047793
UDI-Public(01)05415067047793(10)S000112410(17)270910
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K230286
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDM5000
Device Lot NumberS000112410
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/13/2026
Supplement Date Received by Manufacturer09/04/2026
Initial Report FDA Received Date09/02/2026
Supplement Report FDA Received Date09/09/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/20/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-