D2a: common device name: percutaneous cardiac ablation catheter for treatment of atrial fibrillation with irreversible electroporation; reported here as the common device name exceeded the character limit for designated field.Detailed product information was not provided to bsc.We completed a good faith effort to obtain the information, but it was unable to be obtained.Because the product is unknown, we are unable to provide the unique identifier (udi) and other specific product information.Investigation summary
with all the available information, boston scientific is unable to confirm cause of the reported clinical observation of a foreign material present in device.The device has not been received for analysis; therefore, a technical analysis of the complaint device could not be completed.Device history record review
it was confirmed this device met specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Risk review:
a risk review performed for the farawave device confirmed that the event of foreign material present in device is a known event defined in the products risk management documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Labeling review:
based on the information provided, there is no evidence that the device was used in a manner inconsistent with the labelled indications/instructions for use (ifu).Investigation conclusion
boston scientific has assigned an investigation conclusion code of cause not established and supporting evidence that came to this conclusion.Boston scientifics investigation findings were unable to establish a clear conclusion about the cause of the reported event.The device met all requirements prior to being approved for final distribution/sale and there was no evidence to suggest that the instructions for use (ifu) was not followed.
|