• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. Bard Inc. (Covington) -1018233 Bardex® Lubri-Sil® I.C. All-Silicone Foley Catheter

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. Bard Inc. (Covington) -1018233 Bardex® Lubri-Sil® I.C. All-Silicone Foley Catheter Back to Search Results
Catalog Number 1758SI14
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/02/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during surgery, shortly after starting urinary leakage was seen.No medical intervention was reported.Per follow up information received via ibc on 15jul2025, stated that the actual event for which urinary incontinence happened was unknown and no patient or user was harmed.
 
Event or Problem Description
It was reported that during surgery, shortly after starting urinary leakage was seen.No medical intervention was reported.Per follow up information received via ibc on (b)(6)2025, stated that the actual event for which urinary incontinence happened was unknown and no patient or user was harmed.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Additional Manufacturer Narrative
Evaluation determined the product did not meet specifications.Capa (b)(4)related issues were identified.Received (b)(4) photo samples.The photo sample was returned and could not be evaluated for the reported failure.Visual evaluation of the returned sample noted one opened (with original packaging), used drainage bag with an inlet tubing, sample port connector, catheter, tes.Visual inspection of the sample shows no visual defects present.The reported failure could not be evaluated because it is unknown how the product interacted with the patient's anatomy; therefore, this investigation is considered inconclusive.Inflated the catheter balloon using in-house luer lock syringe with 10 ml methylene blue solution (3 drops 1% aq methylene blue per 100ml distilled water) and the catheter rested with no leaks noted.Deflated balloon passively with no cuffing or leaks noted.Attempted to flush the drainage lumen with blue methyl solution and d.I.Water and it immediately leaked out of a 0.038" pinhole found on the drainage funnel where the silicon mold meets.Further testing noted flushed the sample port and filled the drainage bag with 10 ml methylene blue solution (3 drops 1% aq methylene blue per 100ml distilled water) and no leaks observed.The labeling is found to be adequate.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.Corrections made to tab(s) d and h.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Bardex® Lubri-Sil® I.C. All-Silicone Foley Catheter
Common Device Name
Silicone Foley Catheter
Manufacturer (Section D)
C.R. Bard Inc. (Covington) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. Bard Inc. (Covington) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
+1(770)784-6100
MDR Report Key50578065
Report Number1018233-2026-05684
Device Sequence Number18569008
Product Code MJC
UDI-Device Identifier00801741025211
UDI-Public(01)00801741025211(11)240926(17)290925(10)NGJT2045
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
K984136
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number1758SI14
Device Lot NumberNGJT2045
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/25/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer09/04/2026
Initial Report FDA Received Date09/03/2026
Supplement Report FDA Received Date09/11/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/26/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-