| Model Number |
9738 |
| Medical Device Problem Code |
Sensing Intermittently (1558)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported: "intermittent readings".There was no patient impact or consequence reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: the returned device was evaluated.The device was found to have recessed pins within the patient monitoring cable-connector port.This causes the device to display a "replace sensor" error message when attempting to obtain measurements.The customer's complaint was replicated in testing.Health effect impact code: no adverse event; no patient impact.
|
| |
|
Search Alerts/Recalls
|