Model Number 97810
Medical Device Problem Codes Charging Problem (2892); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/29/2026
Type of Reportable Event Malfunction
Event or Problem Description
Information was received from a patient who was implanted with an implantable neurostimulator (ins) for gastrointestinal/pelvic floor therapy.It was reported that they are having trouble charging the recharger.All three lights are blinking green.It was noted in the repair request that there was scrolling lights on the recharger.The issue started this week.Walked caller through resetting the recharging device.The troubleshooting steps that were taken on the call did not resolve the issue.Agent asked if they store the recharger in the dock when not using it.They said they try to.Sent request to repair to replace the wireless recharger.
 
Additional Manufacturer Narrative
Continuation of d10: product id wr9220 lot # serial # (b)(6); implanted: explanted: product type recharger section d information references the main component of the system.Other relevant device(s) are; product id: wr9220, serial/lot #: (b)(6), ubd; udi#: h3: analysis of the returned wireless recharger (s/n: (b)(6) revealed that the patient's complaint was confirmed.The device led's scroll continuously and the device is unresponsive.The flash read/write protection bits have become set.No visual anomalies were observed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
InterStim
Common Device Name
STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCON
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
+1(763)526-5677
MDR Report Key50579111
Report Number3004209178-2026-14897
Device Sequence Number10195969
Product Code EZW
UDI-Device Identifier00763000203849
UDI-Public(01)00763000203849(17)211014
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P970004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Notification,Recall
Type of Report Initial
Report Date (Section B) 09/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/14/2021
Device Model Number97810
Device Catalogue Number97810
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date09/03/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/16/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0293-2022
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
"See H11...."
Patient Age78 YR