Correction field: d5, d10, e2, g2, g7
updated fields: b4, d9 (return to manufacture date), g3, g6, h2, h6(type of investigation, investigation findings, investigation conclusions, component code), h11
a biomed tech was dispatched to evaluate the unit.The biomed tech reported that failed the battery run time test at 44 min.The run time spec is 135 min.The biomed tech replaced the batteries 2x.The unit passed all functional and safety tests per factory specifications and the unit was returned to the customer.The defective battery was received for evaluation by the getinge failure analysis and testing dept.The following investigation was performed by carl famulare, technician of the maquet failure analysis and testing dept.The failure analysis and testing dept.Received part battery sealed lead acid 12v with a reported unit failure of short battery run time, not holding full charge.Please note this part number is a quantity of two batteries.The failure analysis and testing dept.Conducted a visual inspection and found the parts to be in good condition.The failure analysis and testing dept.Installed battery sealed lead acid 12v into the cs300 test fixture and tested the to factory specifications per the cs300 service manual part
the fat dept.Was able to replicate the complaint of the short battery run time 44 minutes.The factory specification for battery run time is 135 minutes.The batteries failed testing.Retaining the batteries in the fat dept.Based on the fse investigation and the investigation of the returned component - the effect of battery will not charge or battery short run time on the cs300 is confirmed.
|