Datascope Corp. CARDIOSAVE HYBRID, TYPE G PLUG; System, Balloon, Intra-Aortic And Control
|
Back to Search Results |
|
| Model Number |
0998-00-0800-52 |
| Medical Device Problem Codes |
Failure to Power Up (1476); Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/13/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by the customer that during installation, the cardiosave intra-aortic balloon pump (iabp) failed all manifold checks on pim test.There was no patient involved.No harm was reported.A getinge field service engineer (fse) was dispatched to evaluate the unit.The fse reported that pim needs replacement.After pim was replaced, the original fault was resolved but device failed electrical safety test.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated fields: h2, h11, g6, g3, b5, h6 (problem code, component codes) e2, e3, g2.Upon completion of the investigation into this event a follow up report will be submitted.
|
| |
|