| Model Number |
TOPS 30M |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
08/04/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The company was informed about a revision case of the tops system in the us after more than 5 years of implantation.The original procedure was performed on (b)(6), 2021.The patient was treated with tops at l4-l5 as part of the us ide clinical study.Patient experienced initial relief of pain symptoms after surgery but by month 3 the back pain had returned and remained until her completion of the clinical study.The device remained functional for the full duration.The tops system was removed as part of the revision surgery and fusion was performed.During the revision procedure, the surgeon noted that the patient's disc had degenerated and was not moving in flexion/extension.The surgeon had to convert to fusion because the patient was experiencing back pain.
|
| |
|
Additional Manufacturer Narrative
|
|
Review of the manufacturing records of the involved lot was performed and indicated that it was manufactured in accordance with company specifications without deviations.Explant was returned for investigation and examined.The explant analysis identified wear of the internal parts of the implant.Review of x-rays from implantation indicated malplacement of the patient on the operating table, segmental reduction, and a mispositioned tops implant which may have collectively contributed to wear.The rate of revisions for the tops system is lower than the reported rates in the literature for similar systems.
|
| |
|