| Model Number |
LifeVest WCD 5100 System |
| Medical Device Problem Code |
Charging Problem (2892)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A us distributor reported that a patient's battery charger was unable to charge a battery pack.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation of battery charger has been completed.The reported problem (battery charger problem) has been confirmed.Upon investigation the battery charger was unable to charge a battery pack.The cause for the failure was isolated to a defective output plug.The root cause for the defective output plug could not be positively identified.No adverse event resulted from the damaged charger.
|
| |
|