| Model Number |
439688 |
| Medical Device Problem Code |
Extracardiac stimulation (1463)
|
| Health Effect - Clinical Codes |
Device Overstimulation of Tissue (1991); Heart Block (4444)
|
| Date of Event |
03/31/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the left ventricular (lv) lead patient was feeling a thumping in their abdomen while laying on their left side.The patient was stated to have underlying rhythm of complete heart block.Reprogramming was performed and the thumping in the patient abdomen ceased. the patient is a participant in a clinical study.The lead remains in use. no further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|