The customer reported that during 4 circumcisions, the blade cut the skin below where the clamp is placed, and the reporter stated there is a sharp edge that is cutting the patient.A sample was received for investigation; visual and functional inspection was conducted, the simulated grip strength test was performed with a latex glove, and the item was measured with an opening of 2.3mm, which meets all the requirements and specifications.Based on the complaint details, sample evaluation and investigation, the complaint cannot be confirmed.After a functional inspection it was found that the item functions properly and meets all the specification.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|